Chinese Journal of Radiation Oncology ; (6): 462-465, 2015.
Article in Chinese | WPRIM | ID: wpr-467354

ABSTRACT

Objective To study the real?time dose verification with 2D array ion chamber array in volumetric modulated arc therapy ( VMAT) with a 2D array ion chamber array. Methods The 2D ion chamber array was fixed on the panel of electronic portal imaging device (EPID). Source?detector distance (SDD) was 140 cm. 8 mm RW3 solid water was added to the 2D array to improve the signal noise ratio. Patient plans for esophageal, prostate and liver cancers were selected to be delivered on the cylindrical Cheese phantom 5 times in order to validate the reproducibility of doses. Real?time patient transit dose measurements were performed at each fraction. Dose distributions were evaluated using gamma index criteria of 3 mm DTA and 3% dose difference referred to the first time result. Results The gamma index pass rate in the Cheese phantom were about 98%;the gamma index pass rate for esophageal, prostate and liver cancer patient were about 92%, 92% and 94%, respectively. Gamma pass rate for all single fraction were more than 90%. Conclusions The 2D array is capable of monitoring the real time transit doses during VMAT delivery. It is helpful to improve the treatment accuracy.